Abstract :
PERBANDINGAN METODE GRIES DAN METODE SULPANILAT
DALAM PEMERIKSAAN NITRIT PADA SOSIS MENGGUNAKAN
SPEKTROFOTOMETER UV-VIS
Maulana Muhammad, Korry Novitriani, Dewi Kania
(Program Studi D-III Analis Kesehatan/TLM, Universitas Bakti Tunas Husada Tasikmalaya)
Abstrak
Nitrit sering digunakan dalam pengawetan daging untuk meningkatkan warna dan
menghambat pertumbuhan mikroorganisme seperti Clostridium botulinum.
Namun, konsumsi berlebihan natrium nitrit berpotensi menimbulkan risiko
kesehatan, termasuk kerusakan jantung dan kanker. Oleh karena itu, penting untuk
mengukur kadar nitrit secara akurat dan efisien dalam produk daging. Penelitian ini
bertujuan membandingkan efektivitas dua metode analisis yang lazim, yaitu metode
Griess dan metode sulpanilat, dalam mengukur kadar nitrit pada sosis dengan
menggunakan spektrofotometer UV-Vis. Hasil pengujian menunjukkan bahwa
kedua metode tersebut menghasilkan data yang kurang memuaskan. Validasi
metode mengungkapkan bahve presisi dan akurasi kedua metode ini tidak
memenuhi kriteria keberterimaan. Selain itu, Uji Independent Sample T-Test
menunjukkan tidak adanya perbedaan signifikan antara hasil metode Griess dan
metode Sulpanilat. Temuan ini menunjukkan perlunya validasi lebih lanjut untuk
memastikan keakuratan dan keandalan metode-metode tersebut dalam penggunaan
rutin.
Kata Kunci : Nitrit, Sosis, Spektrofotometer, Metode Gries, Metode Sulpanilat.
Abstract
Nitrite is often used in meat preservation to enhance color and inhibit the growth
of microorganisms such as Clostridium botulinum. However, excessive
consumption of sodium nitrite poses potential health risks, including heart damage
and cancer. Therefore, it crucial to accurately and efficiently measure nitrite levels
in meat products. This study aims to compare the effectiveness of two common
analytical methods, the Griess method and the sulfanilate method, in measuring
nitrite levels in sausages using a UV-Vis spectrophotometer. The test results
indicated that both methods produced less than satisfactory data. Method
validation revealed that the precision and accuracy of both methods did not meet
acceptance criteria. Additionally, the Independent Sample T-Test showed no
significant difference between the results of the Griess method and t sulfanilate
method. These findings highlight the need for further validation to ensure the
accuracy and reliability of these methods for routine use.
Keywords: Nitrite, Sausage, Spectrophotometer, Gries Method, Sulpanilate methode